Blood Gas Electrolyte Analyzer HK-2003-B
Item #: AM-COR-0323-ZP
Model: HK-2003-B
Regulatory Status: Approved by Regulations on the Supervision and Administration of Medical Devices.
Sample form: serum, plasma, whole blood, body fluid, urine, etc.
Sample dosage: 60ul ~
Measurement time: ≤45 seconds
Measurement speed: ≤25 seconds
Working temperature: + 5 ℃ ~ + 35 ℃ Storage temperature: -10 ℃ ~ + 50 ℃
Working humidity: ≤85% Storage humidity: ≤95%
Power: ≤60VA
Power supply: AC220 ± 22V 50Hz ± lHz
Application: Medical Use
■ Automatic sampling, automatic calibration, automatic calibration, automatic cleaning, automatic display of results and printing functions.
■ With automatic potential tracking technology, one point calibration, two point calibration, correction of electrode potential drift,
Improve measurement precision.
■ Equipped with two parameters of slope and intercept to ensure the linearity and accuracy of the measurement results.
■ Adopt advanced direct injection sampling and flushing system to avoid clogging and cross-contamination.
■ K +, Na +, Cl-, Ca ++, ACa, TCa, PH, TCO2, AG can be measured simultaneously with a single injection
(Where ACa, TCa, AG are calculated items) parameters.
■ Backlit LCD liquid crystal display, large screen, full Chinese display, built-in thermal Chinese high-speed printer,
Can print sample values, comprehensive reports, and automatically record.
■ Human-machine dialogue, easy to operate. Print head life ≥ 50,000 sample results.
■ Original full PVC active carrier electrode technology, long life, not affected by temperature, no regeneration solution required.
■ The micro-pressure non-contact silicon pressure sensor imported from the United States has high sensitivity and a service life of more than 10 years.
■ Real-time multi-tasking system using the latest embedded processor.
■ A real-time monitoring and tracking system is provided to ensure that the measurement results are accurate and reliable.
■ Intelligent monitoring sensor, can accurately detect the presence of air bubbles in the sample, waste liquid overflow, standard solution
Exhausted, and can automatically alarm.
Rhenium and ISE are measured separately using the same channel, the flow path is simplified, and the sample volume is small.
■ The sealed area of the syringe needle is fully sealed to avoid spilling reagents and causing waste.
■ Built-in three levels of quality control for more accurate calibration.
■ Store all QC data in the last 90 days.
■ Can store more than 10,000 sample results, both by time and by number.
■ Equipped with RS-232 serial interface, compatible with any type of management system and software.
■ Low price of instruments and consumables, minimizing the use cost.
■ Complete and efficient after-sales service network, visit users regularly, eliminating all worries.
VG55 Non-invasive Ventilator, Aeonmed, povide ICU-level non-...
Add to cartAntu fully automatic blood culture system uses advanced non-...
Add to cartAs a professional supplier of medical devices, STEMart-PPE is good at listening, digging and uncovering the unique demands of our customers and committed to providing tailored solutions and products across the continuum of care. Through the power of science, we bring improvement to large healthcare systems, independent physician practices and daily life of home health patients and their families.